Measure sample dimensional changes under conditions of controlled temperature, time, force and atmosphere.

CAPABILITY:

Temperature range -150 to 1000 ˚C

Temperature precision +/- 1 ˚C

Furnace cool down time ≤ 10 min from 600 ˚C to 50 ˚C

Vertical furnace software control

Maximum sample size  solid  26 mm (L) x 10 mm (d)

Maximum sample size film/fiber (static) 26 mm (L) x 1.0 mm (T) x 4.7 mm (W)

Measurement precision +/- 1%

Sensitivity 15 nm

Displacement resolution ≤ 0.5 nm

Precision LVDT over wide temperature range   -150 to 1000 ˚C

Force range  0.001 to 2 N

Force resolution 0.001N

Non-contact motor provides a controlled friction-free calibrated force

Mass flow control included

Atmosphere Inert, oxidizing or reactive gases